Checkline BM User Manual
Page 2

Electromatic Equipment Co., Inc
-2-
1.0
MEASURING PROCEDURE
1. Place the scale on an even and sturdy work surface.
Place the BM2 in the centre of the scale. Zero the scale
to show 0.00kg.
2. Fill the supplied 13 litre bucket with samples taken from
varying locations in sample storage.
3. Check that the measuring chamber is completely empty.
It is important that no material is left in the measuring
chamber when you turn on the device.
4. Switch on the BM2 by pressing the power button
for
3 sec.
5. As the next step, please do the self calibration. The
word “Reinitialize?” will show up on your display. Accept
by pressing the
button.
6. The self calibration is finished when the display shows
the measuring window.
7. Fill the measuring device with the sample evenly
distributed to ensure reproducible results.
8. Remove material till the next lower weight range is
reached. If only a little missing (e.g. 0,10kg) to the next
higher level, it should be filled up to that stage. The
actual weight is shown on the scale.
Electromatic Equipment Co., Inc
-15-
NOTES