0 troubleshooting – Checkline TMI-300 User Manual

Page 21

background image

19

Overload

200% (F.S.) | warning at 120% of the measuring range

Modes of operation

Standard, right-handed, left-handed

(showing peak-values as Nm, Ncm, Kgcm or Lb-in)

Threshold monitoring (adjustable thresholds and

positive/negative indication by coloured LEDs)

Display type

LCD, 4-digits, 12mm high with side indication 7mm

Update rate
(Standard)

125 msec | 250 msec | 500 msec | 1000 msec (adjustable)

Update rate (Peak)

1 msec

Thresholds mode of
operation

6-colours

LEDs

Data memory

quantity : 1000 measurements; statistics function: single,

average, max, min

Energy supply

Internal, rechargeable NiMH battery | Universal power supply- / battery charger

100…240VDC (50Hz)

Battery durability

8…12 h

Low Batt indicator

yes

Interface

USB

1.1

Hirose

socket

Temperature range
(operation)

0°…40° (max. 85%rF)

Temperature range
(storage)

-10°…50 (dry storage)

Weight

8,0 Kg

12,5 Kg

8,0 Kg

12,5 Kg

8,0 Kg

12,5 Kg

Size (L x B x H)

320 x 227 x 126 mm

Casing material

Aluminium cast, protection cover PS

Type of protection

IP44 (splash-water protection)


The instrument, apart from expendable and wearing parts, is maintenance-free. Depending on the area and
frequency of use, torque meters should be regularly, namely at least once a year, be checked and calibrated. Our
calibration service (e-mail address: [email protected]) includes the instrument`s technical check, calibration –
factory calibration – and, if necessary, adjustment of the measurement range.

12.0 Troubleshooting

Display fades out!

Check the remaining voltage. Plug in the adaptor/charger and check the electricity connection.

Values are depicted with inverted prefix!

Change the mode of operation or the prefix; see Chapter 6.4.

No deletion of selected values!

The instrument deletes stored values sequentially; the lastly measured value is deleted first. Selective
deletion is not possible.

Threshold monitoring, data storage and other functions do not work!

Check your instrument`s model. The TMI-100 model is not equipped with the same functions as the TMI-
200 model.

RS-232C communication is disrupted!

Please check the connection cable`s wiring as well as the feed line`s compliance with the PC-card`s
input assignments.

TMI-300; printing is not possible!