0 troubleshooting – Checkline TMI-300 User Manual
Page 21
19
Overload
200% (F.S.) | warning at 120% of the measuring range
Modes of operation
Standard, right-handed, left-handed
(showing peak-values as Nm, Ncm, Kgcm or Lb-in)
Threshold monitoring (adjustable thresholds and
positive/negative indication by coloured LEDs)
Display type
LCD, 4-digits, 12mm high with side indication 7mm
Update rate
(Standard)
125 msec | 250 msec | 500 msec | 1000 msec (adjustable)
Update rate (Peak)
1 msec
Thresholds mode of
operation
6-colours
LEDs
Data memory
quantity : 1000 measurements; statistics function: single,
average, max, min
Energy supply
Internal, rechargeable NiMH battery | Universal power supply- / battery charger
100…240VDC (50Hz)
Battery durability
8…12 h
Low Batt indicator
yes
Interface
USB
1.1
Hirose
socket
Temperature range
(operation)
0°…40° (max. 85%rF)
Temperature range
(storage)
-10°…50 (dry storage)
Weight
8,0 Kg
12,5 Kg
8,0 Kg
12,5 Kg
8,0 Kg
12,5 Kg
Size (L x B x H)
320 x 227 x 126 mm
Casing material
Aluminium cast, protection cover PS
Type of protection
IP44 (splash-water protection)
The instrument, apart from expendable and wearing parts, is maintenance-free. Depending on the area and
frequency of use, torque meters should be regularly, namely at least once a year, be checked and calibrated. Our
calibration service (e-mail address: [email protected]) includes the instrument`s technical check, calibration –
factory calibration – and, if necessary, adjustment of the measurement range.
12.0 Troubleshooting
Display fades out!
Check the remaining voltage. Plug in the adaptor/charger and check the electricity connection.
Values are depicted with inverted prefix!
Change the mode of operation or the prefix; see Chapter 6.4.
No deletion of selected values!
The instrument deletes stored values sequentially; the lastly measured value is deleted first. Selective
deletion is not possible.
Threshold monitoring, data storage and other functions do not work!
Check your instrument`s model. The TMI-100 model is not equipped with the same functions as the TMI-
200 model.
RS-232C communication is disrupted!
Please check the connection cable`s wiring as well as the feed line`s compliance with the PC-card`s
input assignments.