Checkline CAP-T User Manual
Page 6
- 6 -
7-1E
7 data bits, 1 stop bit, even parity
7-1o
7 data bits, 1 stop bit, odd parity
7-2E
7 data bits, 2 stop bits, even parity
7-2o
7 data bits, 2 stop bits, odd parity
7-2n
7 data bits, 2 stop bits, no parity
8-1E
8 data bits, 1 stop bit, even parity
8-1o
8 data bits, 1 stop bit, odd parity
8-1n
8 data bits, 1 stop bit, no parity
8-2n
8 data bits, 2 stop bits, no parity
Ft F
Full data (numeric + units)
Ft n
Numeric data only
7.2 bcd - Mitutoyo BCD settings sub-menu
bcdd Output
disabled
bcdE Output
enabled
nPOL
No polarity (absolute value)
POL
Data with polarity (+ for CW, - for CCW)
7.3 AoFF - Automatic shutoff settings sub-menu
no Disabled
1
1-minute automatic shutoff
5
5-minute automatic shutoff
10
10-minute automatic shutoff
20
20-minute automatic shutoff
30
30-minute automatic shutoff
7.4 init - Initial (default) settings sub-menu
LBIN
Pound-inch as default unit
KGMM
Kilogram-millimeter as default unit
NCM
Newton-centimeter as default unit
CCW
Real time display at turn on
PEAK CW
Peak clockwise display at turn on
PEAK CCW Peak counterclockwise display at turn on
CAL
Calibration sub-menu. See CALIBRATION section.
- 7 -
8.0 C
ALIBRATION
To properly calibrate the CAP-T, application of a precise torque value equal
to the full capacity of the tester in lbin (regardless of the displayed units) will
be required. The procedure follows:
1. Press and hold the PEAK control while turning on power to the
CAP-T tester.
2. When CAL appears on the display, press POWER three times to select
the calibration mode.
3. At the null prompt, press ZERO.
4. At the SPAn prompt, apply the calibration torque and press POWER.
The display will show uuuu or nnnn if the calibration torque is insufficient
or excessive, respectively.
If this happens, the only way to terminate the calibration mode is by
momentarily disconnecting the battery or connecting the AC adapter to the
tester without plugging the other end into a wall outlet. This will stop the
calibration procedure without making any changes to the previous calibration
data.
5. Successful calibration is indicated by donE on the display.
Press POWER to save the changes and resume normal operation.