- 6 -

7-1E

7 data bits, 1 stop bit, even parity

7-1o

7 data bits, 1 stop bit, odd parity

7-2E

7 data bits, 2 stop bits, even parity

7-2o

7 data bits, 2 stop bits, odd parity

7-2n

7 data bits, 2 stop bits, no parity

8-1E

8 data bits, 1 stop bit, even parity

8-1o

8 data bits, 1 stop bit, odd parity

8-1n

8 data bits, 1 stop bit, no parity

8-2n

8 data bits, 2 stop bits, no parity

Ft F

Full data (numeric + units)

Ft n

Numeric data only

7.2 bcd - Mitutoyo BCD settings sub-menu

bcdd Output

disabled

bcdE Output

enabled

nPOL

No polarity (absolute value)

POL

Data with polarity (+ for CW, - for CCW)

7.3 AoFF - Automatic shutoff settings sub-menu

no Disabled
1

1-minute automatic shutoff

5

5-minute automatic shutoff

10

10-minute automatic shutoff

20

20-minute automatic shutoff

30

30-minute automatic shutoff

7.4 init - Initial (default) settings sub-menu

LBIN

Pound-inch as default unit

KGMM

Kilogram-millimeter as default unit

NCM

Newton-centimeter as default unit

CCW

Real time display at turn on

PEAK CW

Peak clockwise display at turn on

PEAK CCW Peak counterclockwise display at turn on
CAL

Calibration sub-menu. See CALIBRATION section.

- 7 -

8.0 C

ALIBRATION

To properly calibrate the CAP-T, application of a precise torque value equal
to the full capacity of the tester in lbin (regardless of the displayed units) will
be required. The procedure follows:

1. Press and hold the PEAK control while turning on power to the

CAP-T tester.

2. When CAL appears on the display, press POWER three times to select

the calibration mode.

3. At the null prompt, press ZERO.

4. At the SPAn prompt, apply the calibration torque and press POWER.

The display will show uuuu or nnnn if the calibration torque is insufficient
or excessive, respectively.

If this happens, the only way to terminate the calibration mode is by
momentarily disconnecting the battery or connecting the AC adapter to the
tester without plugging the other end into a wall outlet. This will stop the
calibration procedure without making any changes to the previous calibration
data.

5. Successful calibration is indicated by donE on the display.

Press POWER to save the changes and resume normal operation.