beautypg.com

Checkline CAP-T User Manual

Page 5

background image

- 5 -

6.0 C

HANGING

U

NITS OF

M

EASURE

1. Press and Hold the PEAK key and at the same time. Press the POWER

key. (The Display will show "Aoff"). Then release all keys.

2. Press the PEAK key one (1) or more times until the display shows "init"

3. Press the POWER key one time. (The currently selected unit of measure will

flash on/off on the right side of the display)

4. Press the PEAK key to cycle thru the choices of units. Each time the PEAK

key is pressed the new units selection will be shown on the right side of the
display. (The desired unit of measure should be flashing before you continue
to step #5.)

5. Press the POWER key two (2) times to store the new units selections. Display

will show "donE".

6. Press POWER again to return to the measurement mode with the new unit

selection.

7.0 C

ONFIGURATION

CAP-T torque testers have several features with programmable options
allowing many user-specified choices. To enter the configuration menu,
perform the following:

1. Turn off the tester
2. Press and hold PEAK
3. Turn on the tester
4. Release PEAK

The version number of the internal software will be displayed for a short time
followed by either 'AoFF' for a standard CT or '232' if equipped with CTRS
communication option. The following secondary functions of keys are used
during the configuration process:

POWER

Used to select a menu choice

PEAK

Used to step through menu choices

7.1 232 - RS-232 settings sub-menu

232d Output

Disabled

232E Output

Enabled

300 300

baud

600 600

baud

1200 1200

baud

2400 2400

baud

4800 4800

baud

9600 9600

baud

- 8 -

9.0 O

PERATION

Place the sample between the posts or jaws of
the tester, and tighten,using the knob. Posts can
be placed in any of the holes on the sliders.

Ensure that the sample is gripped firmly,

however, excessively tightening thin- walled

closures may result in sample deformation,
possibly affecting test results. When the sample
is secured, gradually exert torque by hand.

NOTE: Sudden torque application may produce

incorrect readings.

The default mode of operation is the

normal (real time) mode.

If the peak readings are to be observed as they occur,
then the mode of operation can be changed by press-
ing PEAK until the desired mode (CW PEAK or
CCW PEAK) appears on the display.

This action affects only the display. The peak
readings are captured automatically and can be
recalled by pressing PEAK. The peak readings
may be cleared from the memory by either pressing
ZERO or shutting off power

9.1 Changing Operating Modes and Measuring Units

The displayed units of measurement and the default mode of operation
(peak or normal) can be changed by entering the setup mode as described in
section 6.0, pressing POWER at the init prompt and selecting the desired
settings using the PEAK and the POWER keys.

9.2 Auto

Shut-off

The CAP-T is equipped with an automatic shutdown feature and it will shut
off after approximately 30 minutes of inactivity (readings do not change by
more than ±10 counts and no keys are pressed). The entire display will flash
for 5-7 seconds as a warning of the imminent shutdown.

To change the default setting of 30 minutes, hold PEAK while turning on
the gauge. Press PEAK repeatedly until 'AOFF' appears. Press POWER to
select this function. The current setting will flash on the display. Use PEAK
to scroll through the displayed choices and POWER to select.

Knob