Checkline MX-5DL User Manual
Page 26

Dakota Ultrasonics
22
Differential Mode
In the Quality Control environment, it is sometimes necessary to know
the difference between a nominal (target) thickness value and an actual
thickness value. This feature is also included in the MX-5 DL. With the
Differential Mode enabled, the MX-5 DL will display the positive or negative
difference from an entered nominal value. The following steps outline the
procedure for setting up this feature:
Differential Mode
1) Press ON/OFF key to power up the MX-5 DL.
2) Press the MODE key to activate features and settings.
3) Press the UP or DOWN arrow keys to scroll to dIFF.
4) Press the SEND key to toggle the staus of the differential mode
on/off.
5) Status ON - A nominal value will be displayed with the units IN/MM
flashing.
6) Press the UP or DOWN arrow keys to scroll to the desired nominal
thickness value.
- Series-2 (16 pages)
- Series-3 (22 pages)
- Series-4 (26 pages)
- Series-5 (34 pages)
- FGV-XY (19 pages)
- DS2 (6 pages)
- FGE (8 pages)
- Series-5I (32 pages)
- Series-3I (24 pages)
- MG (2 pages)
- FGE-XY (6 pages)
- FB (4 pages)
- ES10 (4 pages)
- FGS-100H (2 pages)
- FGS-250W (2 pages)
- TSB100 (10 pages)
- ES30 (6 pages)
- ES05 (4 pages)
- TSA750 (10 pages)
- TSF (6 pages)
- ESM301 (27 pages)
- MX2 (8 pages)
- FGS-100PV (20 pages)
- FGS-220VC (33 pages)
- MX-500 (4 pages)
- ESMH (6 pages)
- FGS-100PX (20 pages)
- ZP (10 pages)
- WT-100 (2 pages)
- AWS-4050 (14 pages)
- TI-25LT (9 pages)
- TI-25S (11 pages)
- TI-25M (13 pages)
- TI-25DL (19 pages)
- TI-25M-MMX (15 pages)
- TI-25DL-MMX (17 pages)
- TI-007 (9 pages)
- TI-007DL (11 pages)
- TI-CMX (29 pages)
- TI-CMXDL (35 pages)
- TI-CMXDLP (54 pages)
- TI-MVX (42 pages)
- TI-UMX2 (29 pages)
- TI-25P (11 pages)
- TI-44N (11 pages)