Series 4 Digital Force Gauges
User’s Guide
15
9.6 I/O Connector Pin Diagram (female)
Pin No.
Description
Input / Output
1
Signal Ground
---
2
Tension Overload
Output
3
RS-232 Receive
Input
4
RS-232 Transmit
Output
5
+12V DC
Output
6
Analog Output
Output
7
Compression Overload
Output
8
Mitutoyo Clock
Output Bit 2
Output
9
Mitutoyo Data
Output Bit 0
Output
10
Mitutoyo Request
Input Bit 3
Input
11
“Under” Set Point
Output
12
“Over” Set Point
Output
13
“Within” Set Point
Output
14
External Trigger
Input
15
Mitutoyo Ready
Output Bit 1
Output
10 CALIBRATION
10.1 Initial Physical Setup
The gauge should be mounted vertically to a test stand or fixture rugged enough to withstand a load
equal to the full capacity of the instrument. Certified deadweights or master load cells should be used,
along with appropriate mounting brackets and fixtures. Caution should be taken while handling such
equipment.
10.2 Calibration Procedure
1. Select Calibration from the menu. The display will appear as follows:
CALIBRATION
To invert the
display, press the
DIRECTION button.
THEN PRESS ENTER