Data key, Rs232/usb output, Mitutoyo output – Checkline Series-4 User Manual
Page 15: Memory storage, Mitutoyo bcd, Disabled, Ena w/o polarity, Ena w/polarity
![background image](/manuals/656223/15/background.png)
Series 4 Digital Force Gauges
User’s Guide
14
Any detected errors are reported back by means of the following error codes:
*10 Illegal command
*11 Not applicable
*12 Invalid specifier
*22 Value too large
9.3 Mitutoyo BCD settings
This output is useful for connection to data collectors, printers, multiplexers, or any other device capable
of accepting Mitutoyo BCD data. Individual data points may be transmitted by pressing DATA or by
requesting it from the Mitutoyo communication device (if available). To enable Mitutoyo output, select the
desired format – either with polarity or without polarity. The screen appears as follows:
9.4 Analog Output
This output can be used for chart recorders, oscilloscopes, data acquisition systems, or any other
compatible devices with analog inputs. The output produces ±1 volt at full scale of the instrument. The
polarity of the signal is positive for compression and negative for tension.
9.5 DATA Key Functions
The DATA key can be configured to perform several functions. To configure the DATA key, select DATA
Key from the menu. The display will appear as follows:
Three options are available:
Selection
Function when pressing DATA
RS232/USB Output
Outputs data via the serial and USB ports
Mitutoyo Output
Outputs data via Mitutoyo (Digimatic) through the serial port
Memory Storage
Stores a reading to memory (refer to the Memory section for details)
Any combination of the above functions may be selected.
MITUTOYO BCD
*
Disabled
Ena w/o Polarity
Ena w/Polarity
DATA KEY
*
RS232/USB Output
Mitutoyo Output
Memory Storage