Checkline Series-3 User Manual
Page 16
Series 3 Digital Force Gauges
User’s Guide
15
Press ENTER.
12. The display will appear as follows:
Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees. Press
DIRECTION to invert the display. Then attach weight fixtures. The following screens will step
through the same procedure as with the compression direction. Proceed in the same manner.
13. At the completion of the tension calibration, the display will appear as follows:
To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.
14. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
CALIBRATION
To invert the
display, press the
DIRECTION key.
THEN PRESS ENTER
CALIBRATION
Units must be gF.
PLEASE TRY AGAIN
PRESS ENTER
CALIBRATION
COMPLETE
SAVE & EXIT
EXIT W/O SAVING
CALIBRATION
COMPRESSION COMPLETE
REVERSE DIRECTION
FOR TENSION
Attach necessary
weight fixtures.
THEN PRESS ENTER