beautypg.com

Checkline Series-2 User Manual

Page 15

background image

Series 2 Digital Force Gauges

User’s Guide


14

9.4 Initial Mode
This section is used to configure the initial mode upon powering on the gauge. To access this parameter,
select Initial Mode from the menu. The screen will appear as follows:


The default value is Real Time.

9.5 Information / Welcome Screen
The following screen is displayed at power up and can be accessed at any time by selecting Information
from the menu:


10 SPECIFICATIONS

10.1 General

Accuracy:

±0.5% of full scale

Sampling rate:

500 Hz

Power:

AC or 9V non-rechargeable battery, with multi-step low battery indicator

Battery life:

Backlight on: up to 24 hours of continuous use

Backlight off: up to 92 hours of continuous use

Safe overload:

150% of full scale (display shows “OVER” at 110% and above)

Weight:

0.7 lb [0.33 kg]

Included items:

9V battery, quick-start guide, resource CD, certificate of conformance

Load cell deflection:

0.010 in [0.25 mm]

Environmental
requirements:

40 - 100°F, max. 96% humidity, non-condensating

Warranty:

3 years (see individual statement for further details)


Digital Force Gauge

Series 2

Model No: M2-50
Serial No: 1234567
Version: 1.0
(c) Mark-10 Corp.

INITIAL MODE

*

Real Time

Peak Compression
Peak Tension