8 .2 device fuses, Warning – BUCHI SpeedDigester K-436 User Manual
Page 48
![background image](/manuals/656080/48/background.png)
8 Troubleshooting
48
K-425 / K-436 Operation Manual, Version A
8 .2
Device fuses
To change a defective fuse proceed as follows
Switch off the SpeedDigester and unplug the power cord.
•
To access the glass fuses on the rear side use a flat screwdriver blade.
•
➥ Turn it approx. 5 turns anticlockwise to free the insert.
➥ Take out the insert together with the fuse.
Replace the defective fuse (see technical data for fuse type!)
•
Reconnect the SpeedDigester to the mains.
•
!
WARNING
Death or serious burns by electric current.
Switch off the instrument
•
Disconnect the power cord and prevent unintentional restart before removing the fuse holder
•
Do not touch the fuse or the fuse holder with wet hands
•
Exchange defective fuses only with the original type
•
T 10A L 250 V
Note
Device fuses may blow occasionally due to voltage peaks under high system load. When fuses burn
frequently, inform your customer service!