Warning, Caution – BUCHI SpeedDigester K-436 User Manual
Page 28

6 Operation
28
K-425 / K-436 Operation Manual, Version A
!
WARNING
Death or serious poisoning by contact or incorporation of harmful substances at use.
Before operation, check the instrument for correct assembling
•
Close unused suction module positions with glass caps
•
Rinse suction modules, hoses and connectors before use
•
Before operation inspect sealings and tubes for good condition and tightness
•
Exchange worn out or defective parts immediately
•
Only operate the instrument in a fume hood
•
Directly withdraw released gases and gaseous substances by a Scrubber or a water jet pump
•
Dispose of condensate in tubings and glassware safely after use
•
!
CAUTION
Risk of minor or moderate cuts when handling damaged glass parts.
Handle glass parts with care
•
Do not heat up empty sample tubes in the device
•
Visually inspect every glass part before mounting
•
Exchange damaged glass parts immediately
•
Do not touch cracks or bits of broken glass with bare hands
•
6 .2 .1
Kjeldahl digestion —preparing sample tubes
During the preparation process, handling with strong acid and other dangerous substances is inevi-
table. Thus, all steps must be performed under safe laboratory conditions!
Prerequisites
All sample tubes must be clean and free of damages.
•
To achieve a reliable and reproducible digestion result the quality of sample material and additives
•
(e.g. acid and catalyst) must be free of impurities and of adequate amount.
Filling of a sample tube
Take the sample tubes sequentially and weigh in sample material.
•
Add calatyst to the samples (e.g. Kjeldahl tablets, order no. 028765).
Carefully(!)
•
add sulfuric acid (
≈ 98 %) of sufficient amount to the samples.
➡ The sample tubes can now be installed in a rack.
Note
Note down the sample weight and position to keep the probe processing and its analysis repro-
•
ducible. All racks are marked on top with a "1" for the first sample position.
The rack(s) can be used as a temporary storage for empty tubes during the preparation process.
•
When adding the sulfuric acid, rotate the tube to wash down any sticky sample material to the
•
bottom and to coat the inside of the tube.