6 Operation
29
K-425 / K-436 Operation Manual, Version A
6 .2 .2
Peroxide digestion —preparing sample tubes
During the preparation process, handling with peroxides and other dangerous substances is inevi-
table. Peroxides are strong oxidants and can form explosives! All steps must be performed under safe
laboratory conditions.
!
WARNING
Death or serious injuries by toxic and explosive peroxides.
Use only damage free sample tubes
•
Exchange worn out or defective parts immediately
•
Handle peroxides in a fume hood
•
Do not bring peroxides into contact with highly combustible material
•
Prerequisites
All sample tubes must be clean and free of damages.
•
To achieve a reliable and reproducible digestion result the quality of sample material and the
•
peroxide must be free of impurities and of adequate amount.
Filling of a sample tube
Take the sample tubes sequentially and weigh in sample material.
•
Carefully(!)
•
add diluted sulfuric acid (e.g. 69%) of sufficient amount to the samples.
Carefully(!)
•
add peroxide (e.g. H
2
O
2
30%) of sufficient amount to the samples.
➡ The sample tubes can now be installed in a rack.
Note
Note down the sample weight and position to keep the probe processing and its analysis repro-
•
ducible. All racks are marked on top with a "1" for the first sample position.
The rack(s) can be used as a temporary storage for empty tubes during the preparation process.
•
When adding the sulfuric acid and peroxide, rotate the tube to wash down any sticky sample
•
material to the bottom and to coat the inside of the tube.
6 .2 .3
Installing a suction module and sample tubes (300 ml)
To withdraw hazardous fumes and vapors a suction module must be installed.
Prerequisites:
All hoses and connectors must be unclogged.
Installation steps:
For installation and handling all parts must be < 40 °C!
Install the plastic adapter piece to the suction module
•
1.
Mount the hose connector at the EPDM hose
•
2 of the suction module (i.e. Scrubber B-414 or
water jet pump).
The connector at the EPDM hose can be stored temporarily at the magnetic bracket at the rear
•
side of the SpeedDigester.