7 Maintenance and repairs

43

K-425 / K-436 Operation Manual, Version A

7 .2 .1

Breakage of glassware inside a housing chamber

Under rare conditions charged sample tubes or other glassware might break inside a housing
chamber. In such a case strictly follow the cleaning instructions below!

Cleaning under safe conditions

Switch off the SpeedDigester and unplug the power cord.

➡ Let the system cool down completely!

Carefully remove the installed rack(s).

Wear safety gloves to dispose of the defective glassware of the rack.

Use a long forceps to remove residual bits of glass off the housing chamber(s).

Use a long forceps to dab residual liquid and other components off the chamber(s).

Wear cut- and acid-safe gloves to wipe out the chamber(s) with wet damp cloth.

Check the surface below the instrument and clean if necessary.

Wait for the system to completely dry before reconnecting the device to mains!

7 .2 .2

Liquid ingress into the instrument

In case of liquid ingress into the housing the instrument is not safe to use anymore. Follow the instruc-
tions below to safely bring the instrument into operable condition again.

Switch off the SpeedDigester and unplug the power cord.

Use a long forceps to dab liquid off the outer housing parts. Wear safety gloves!

Carefully remove rack(s) if installed.

➡ Let the system cool down completely!

Use a long forceps to dab residual liquid off the chamber(s).

Wear cut- and acid-safe gloves to wipe out the chamber(s) with dry cloth.

Use a long forceps to clean the chamber(s) with a wet damp cloth.

➡ Call service! Do not reconnect the system to mains!

!

WARNING

Death or serious burns by electric current.

Switch off the instrument

Disconnect the power cord and prevent unintentional restart

All parts must be completely dried before the system can be checked by a service technician!
An electrical safety test and a functional test must be performed before the system can be used again.