5 Putting into operation

21

K-425 / K-436 Operation Manual, Version A

5

Putting into operation

This chapter describes how the instrument has to be installed. It also gives instructions for the initial
startup.

Note
Inspect the instrument for damage during unpacking. If necessary, prepare a status report immedi-
ately to inform the postal company, railway company or transportation company. Keep the original
packaging for future transportation.

5 .1

Installation site

Put the instrument into a fume hood onto a clean, stable and horizontal surface. Consider the
maximum product dimensions and weight. Obtain the environmental conditions as described in
section 3.2, technical data.

Installation prerequisites and installation steps:

The fume hood must be equipped with a heat- and acid-resistant lining.

Do not place any objects on top or below the instrument.

The instrument must be installed with a safety clearance of not less than 5 cm to any other

objects or walls to allow sufficient cooling.
Do not place containers, chemicals or other items behind the instrument.

When using a Scrubber B-414 it must be placed on the

left side of the SpeedDigester.

Note

T o cut the power in case of an emergency by unplugging, the instruments or any other items

must not block the mains plug!

Any

cooling of the SpeedDigester can disturb the digestion process. When switched 'On',

the Scrubber B-414 cooling fan emits from the left side of the housing. Hence the Scrubber B-414
should not be installed on the

right side of the SpeedDigester.

!

DANGER

Death or serious injuries by use in explosive environments.

Do not store or operate the instrument in explosive environments

Do not store chemicals in the vicinity of the device

Operate the device in a fume hood with sufficient ventilation to directly withdraw fumes