5 Putting into operation

22

K-425 / K-436 Operation Manual, Version A

!

WARNING

Risk of death or burns by electric current.

Check for proper grounding before use

Exchange defective cabling instantly

!

CAUTION

Risk of minor or moderate injury by heavy weight of the instrument.

Lift the instrument carefully and avoid over exertion

Do not drop the instrument or its transport box

Place the instrument on a stable, even and vibration-free surface

Keep limbs out of crushing zone

5 .2

System fixation

In earthquake prone regions the instrument must be fixed to the fume hood bench. The penetration
depth of the stainless steel fixation screw must not be more than 10 mm.

Notice

Risk of instrument damage by earthquakes

Safely attach the instrument to the bench in earthquake prone regions

Use acid resistant stainless steel screws

System fixation point
(insert nut)

M4Ч10