5 Putting into operation
22
K-425 / K-436 Operation Manual, Version A
!
WARNING
Risk of death or burns by electric current.
Check for proper grounding before use
•
Exchange defective cabling instantly
•
!
CAUTION
Risk of minor or moderate injury by heavy weight of the instrument.
Lift the instrument carefully and avoid over exertion
•
Do not drop the instrument or its transport box
•
Place the instrument on a stable, even and vibration-free surface
•
Keep limbs out of crushing zone
•
5 .2
System fixation
In earthquake prone regions the instrument must be fixed to the fume hood bench. The penetration
depth of the stainless steel fixation screw must not be more than 10 mm.
Notice
Risk of instrument damage by earthquakes
Safely attach the instrument to the bench in earthquake prone regions
•
Use acid resistant stainless steel screws
•
System fixation point
(insert nut)
M4Ч10
2Ч