4 verification, 5 settings, sysinfo, test, 4 verification 6.5 – BUCHI Melting Point M-565 User Manual
Page 45: Settings, sysinfo, test, 6 .5 settings, sysinfo, test

6 Operation
45
Melting Point M-565 Operation Manual, Version D
6 .4 .4
Verification
To verify the calibration, proceed as follows:
• Measure all standards (BUCHI recommends using the verification set (11055019) but it is also
possible to use your internal standards.
NOTE
The verification is not a guided process.
• When all measuring results were within the required tolerances, click on
Verified. The following
screen appears:
• Enter the following password:
VER. The current date is now indicated under “Last verification”.
6 .5
Settings, SysInfo, Test
Test: Functional test of the instrument. To per-
form the test of several functions follow carefully
the instuction on the screen. If a function test
shows not
ok, please contact the customer
service. The Test protocol is described in sec-
tion 6.5.1.
SysInfo: Instrument gives information regarding
settings and connected devices.
Settings: This menu can be used to change
parameters. The SysInfo protocol is described
in section 6.5.2.