3 creating a method, 4 adjusting parameters during a determination – BUCHI Melting Point M-565 User Manual
Page 40

6 Operation
40
Melting Point M-565 Operation Manual, Version D
73.2
76.4
73.4
76.6
Delay: to ensure equilibrium
between oven temperature
and sample temperature.
During this delay time some
air bubbles turns out of the
boiling point capillary. This
request can be changed in
the setting “Boiling sample
conditioning” (see 6.5).
• To run through the results,
turn the rotary knob up and
down.
Video replay is not available
for boiling point determination.
• Press
End or STOP to
return to the idle screen.
• Press
START to measure
again with the identical
measuring parameters.
6 .3 .3
Creating a method
NOTE
The process of creating a boiling point method is identical that used to create a melting point method
(see 6.2.3).
6 .3 .4
Adjusting parameters during a determination
During preheating or a determination, measuring parameters can be modified. This function is possible
only if enabled by the setting “Param. adj. during meas.” (see 6.5).
a
• Turning the rotary knob
displays the current param-
eter set.
• Adjust the parameters by
pressing Edit. Adjusting the
parameters does not affect
the saved method.
If the temperature gradient
is changed during a mea-
surement, the result shows
CHANGED (a) for the tem-
perature gradient.