5.14 Trouble-shooting: The single platform

Trouble-shooting: the single platform

Observation

Cause

Correction

Display does not light

No power

Connect the unit to the power system.
Check fuses. Check the power system

Display lights only in part

Display unit defective

Contact BÜCHI service agent

Unit does not heat up

Heater not turned on

Press „START“

Unit does not heat up

Heater not turned on

Temperature set-point too low; Setpoint
temperature higher than room temperature.

Unit does not heat up

Heater defective

Contact BÜCHI service agent

Unit does not shake

Drive not activated

Turn speed control clockwise

Unit does not shake

Drive blocked by the adjustment

wrench

Remove the adjustment wrench

Unit does not shake

Drive belt torn

Contact BÜCHI service agent

Unit does not shake

Drive defective

Contact BÜCHI service agent

Vacuum cover is not heated

The 24 V connection socket has not

been connected correctly

Check the plug on the vacuum connection

Vacuum cover is not heatedt

Setpoint temperature too low

Raise the setpoint temperature

Vacuum cover is not heated

Heater defective

Contact BÜCHI service agent

Unit cannot be programmed.

Program mode has not been

activated

Press the „PROGRAM“ button until 1 appears in

the Step display.

Buttons do not react to pressure

Defective contact

Contact BÜCHI service agent

Error message in the display

Mechanical or electronic defect

Contact BÜCHI service agent, refer to Chapter

„Maintenance: Error messages“

Unit shakes but makes noise

when doing so

Vibration elements defective

Inspect vibration elements visually and contact
BÜCHI service agent

Strong vibrations

Settings for unbalance not optimum

Readjust unbalance compensation

Strong vibrations

Rotational speed not properly

adjusted

Reduce speed