Brookfield Temperature Controller, Model 106 User Manual
Page 7
Brookfield Engineering Labs., Inc.
Page 7
Manual No. M/98-206-A0402
specially keyed so that other loads cannot be readily substituted. The power limit supplied by this
connector is 300 watts.
Note:
The current provided to the Thermosel connector is potentially dangerous. Do not insert
or remove the Load plug while power is applied to the Programmable Temperature Con-
troller.
IV. CONTROL KEYS AND DISPLAY PANEL
The front panel of the unit, which includes all user controls and status indicators, is shown in Figure
2:
HEAT
RUN
Programmable Temperature Controller
RUN
STBY
°
F
°
C
PROGRAM
REMOTE
PGM
SET
Figure 2
The digital display is used to display system status, to set up temperature/time programs, to review
temperature/time programs and to present certain messages to the user.
The allowable range for temperature entries is:
0.0
°
C to 300.0
°
C (32.0
°
F to 572.0
°
F), with a minimum increment of 0.1
°
(either scale). The
allowable range for time entries is from 1 to 900 minutes with a minimum increment of 1 minute.
If a proposed temperature value is above the maximum input for that unit, or if a proposed tempera-
ture value is below the minimum input for that unit, an audible sound (beep) will be heard and the
current maximum (or minimum) value will be displayed.
The various buttons and light emitting diodes (LED’s) have the following functions or meanings: