4) The display now reads:

100.00

We have now altered the current program step to correspond to the first entry for Table 7.

5)

Press the

SET

key to accept this value and move to the next program step.

45.0 1

45 is the previously set temperature for the second entry.

1 indicates the second program entry.

Enter the desired temperature for the second entry (140) and press the SET key when the
correct value has been entered.

6) Continue to enter additional steps for the temperature program. The Programmable Temperature

Controller can accept up to 10 entries. The tenth entry is indicated by the integer 9. When you
have entered the seventh and last temperature (indicated by a 6 in the program step display posi-
tion), you must press the SET key four (4) times; once to accept the last program temperature, and
three (3) times to advance the program step counter past the last step (Step 9).

NOTE:

No matter how many steps you have in a program you will always have to press the

SET

key ten (10) times to get to the time input step. This also holds if you simply

want to review a program. If no temperature is in the step, the default temperature is
17.89.

7) After the last press of the SET key in step 6 above you will be presented with the step time input

screen shown here:

23

Use the ARROW keys to change the displayed time (23 minutes in this case) to the hold time of

15

minutes from Table 10.

8) Press the SET key once to accept the hold time interval and to advance to the