beautypg.com

Brookfield Temperature Controller, Model 106 User Manual

Page 16

background image

Brookfield Engineering Labs., Inc.

Page 16

Manual No. M/98-206-A0402

perature Controller will maintain the 160

°

C setting until power is shut off or the

RUN

key is

pressed to stop the program. You may have noticed that this single setpoint program using the
hÏlË

hÏlË

hÏlË

hÏlË

hÏlË option is exactly the same as the Single SetPoint Operation - Non Program Mode except

that we are in the program mode!

V.7 Reviewing and Editing An Existing Program

To review an existing program, press the

PROG

key. The PROGRAM LED will be illuminated.

Simply press the

SET

key to “scroll” through all the various steps of your program. When you are

done, press the

PROG

key once more to exit program review. If you find a program step that is in

error (or simply needs to be changed), advance to that step using the

SET

key and then use the

ARROW

keys to change the step to the correct (or new) temperature value or mode if you are

altering the

hÏlË/s‰Ïp

hÏlË/s‰Ïp

hÏlË/s‰Ïp

hÏlË/s‰Ïp

hÏlË/s‰Ïp option.

V.8 Running A Temperature/Time Program

To run a program, press the

PROG

key (if the PROGRAM LED is already illuminated, you do not

have to press the

PROG

key), and then press the

RUN/STBY

key. The PROGRAM LED and the

RUN LED will now both be illuminated.

When the

RUN/STBY

key is pressed, Programmable Temperature Controller will begin ramping to

that temperature. The HEAT LED will be illuminated when the Programmable Temperature Control-
ler is providing power to the Thermo Container. When the temperature has reached the set point for
that step (+/- 1

°

F or +/- 0.5

°

C), the Programmable Temperature Controller will maintain that tem-

perature for the programmed time period.

While the Programmable Temperature Controller is in the hold time interval for a given setpoint, you
can press the

SET

key twice to obtain the following information:

1) The first press of the SET key will display the current setpoint temperature and the current step

number:

165.03

As shown here, we are holding in step 3 with a setpoint temperature of 165

°

C (or

°

F if that had

been selected as the temperature display units).

2) The second press of the SET key will display the hold time remaining in step 3 before the Pro-

grammable Temperature Controller begins ramping up (or down) to the temperature programmed
in step 4. The time remaining display would appear as follows:

8 ‰

The display shows that there are between 8 and 9 minutes left in the hold time period for this step.
It is important to remember that the Programmable Temperature Controller starts all program
steps at 0, and that all hold times are decremented to 0. This means, for instance, that if 15