Brookfield Temperature Controller, Model 106 User Manual
Page 18
Brookfield Engineering Labs., Inc.
Page 18
Manual No. M/98-206-A0402
perature Controller are formatted as a four digit integer number. Setpoint tempera-
tures being sent should first be multiplied by a factor of ten and expanded to four
digits with leading zeros if necessary. All temperatures (sensed or setpoint) being
read from the Programmable Temperature Controller should be divided by a factor of
ten to obtain the actual temperature.
The operator/user can develop a variety of programs for controlling the Programmable Temperature
Controller via the RS-232C protocol to meet specific needs of the laboratory. The following commu-
nication protocol can be used in any desired combination:
1) To read the current temperature, enter T
{CR}
at the external device.
The response from the Programmable Temperature Controller to the external device will be:
T{tttt}{u}{st}{CR}
where:
tttt
indicates the current temperature multiplied by a factor of ten;
u
indicates the temperature scale (
°
F or
°
C);
st
indicates the present state of the Programmable Temperature
Controller (as described in Table 8 at the end of this section).
A typical response might be
T2345F1,
which means that the temperature is 234.5
°
F, and the con-
troller is presently in the run mode (1).
2) To read the current temperature setpoint, enter
S{CR}
at the external device. The response from
the Programmable Temperature Controller to the computer will be:
S{spspspsp}{u}{st}{CR}
where:
spspspsp
indicates the current setpoint multiplied by a factor of ten;
u
indicates the temperature scale (
°
F or
°
C);
st
indicates the present state of the controller.
A typical response might be
S1004C2,
which means that the current setpoint is 100.4
°
C, and the
controller is in the wait or standby mode (i.e. the Programmable Temperature Controller is not
controlling the Thermo Container).
3) To adjust the temperature setpoint, enter
RS{spspspsp}{u}{CR}
at the external device where
spspspsp
is the setpoint temperature multiplied by ten and
u
indicates the temperature scale in
either
°
F or
°
C.
A typical command sent to the controller might be
RS1004C{CR},
which indicates the new setpoint
should be 100.4
°
C. The generic form of the response from the Programmable Temperature
Controller to the computer will be in the form:
RS{spspspsp}{u}{st}{CR}