Manual No. M13-167-B0614
❏
Viscometer Will Not Return to Zero
✓
Viscometer is not level
• Check with spindle out of the sample.
• Adjust the laboratory stand.
✓
Pivot point or jewel bearing faulty
• Perform an Oscillation Check*
✓
Remove the spindle and turn the motor OFF; select display to % torque mode.
✓
Gently push up on the viscometer coupling.
✓
Turn the coupling until the digital display reads 10-15 on the % display.
✓
Gently let go of the coupling.
✓
Watch the digital display; you should see a “run” of numbers next to the %; the
“run” of numbers
should ultimately stop at 0.0 (+/- 0.1).
If the digital display does not return to ZERO, the unit most likely is in need of
service.
• Perform calibration check (see Appendix F).
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair (see Appendix O).
* This procedure does not apply to instruments with ball bearing suspension (see Section
I.3).
❏
Display Reading Will Not Stabilize
✓
Special characteristic of sample fluid. There is no problem with the viscometer.
• Refer to Appendix C.
✓
Check for erratic spindle rotation.
• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair.
✓ Bent spindle or spindle coupling.
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair.
✓
Temperature fluctuation in sample fluid.
• Use temperature bath for control.