Appendix H - Speed Selection
Brookfield Viscometers offer a variety of speeds to provide for a wide range of viscosity
measurement capabilities. Brookfield has traditionally supplied a defined set of speeds
with specific Torque ranges:
LV
0.3, 0.6, 1.5, 3.0, 6.0, 12, 30, 60
RV
0.5, 1.0, 2.0, 2.5, 4.0, 5.0, 10, 20, 50, 100
The DV2T and earlier DV-II series Brookfield Viscometers offer additional speeds to
enhance measurement capabilities. The DV2T offers speeds from 0.1 - 200 RPM with
two options for speeds selection: numeric key pad for direct entry; Scroll List. The user
can choose from these two entry options by using the Navigation Bar (see Section IV.3:
Global Settings). The Scroll List is the format used with previous versions of the DV-II
viscometer. You may choose the Scroll List to provide the viscometer user with a familiar
method for speed selection. The Scroll List can be customized to limit the available
speeds and to add new speeds to the list of standard speeds.