Brookfield DV2T Extra Viscometer User Manual
Page 20
Brookfield Engineering Labs., Inc.
Page 20
Manual No. M13-167-B0614
Range
Figure II-6
TIP: The Range value is the same as the AutoRange available on earlier Brookfield
viscometer models.
II.7 Out of Range
The DV2T Viscometer will give on screen indications when the measurement is out of range
of the instrument. When the %Torque reading exceeds 100% (over range), the display
of %Torque, Viscosity, and Shear Stress will be EEEE (see Figure II-7). If the %Torque
value is between 0 - 9.9%, the data field label will flash. When the %Torque is below zero
(negative values), the display of Viscosity and Shear Stress will be ----.
TIP: Brookfield recommends collecting data only when the %Torque reading is
between 10 - 100%.
Figure II-7