VIEW CALIBRATION VALUES (F18)

1. While in the Setup mode, scroll to "F 18", then scroll down once using the key to enter View

Calibration menu.

2.

The display will momentarily show "CAL 0" followed by a value. This value is the zero

calibration value and should be recorded in the table below. Press any key to continue.

3. The display will momentarily show "CAL 1" followed by another value. This value is the span

calibration value and should also be recorded in the table below. Press any key to return to
upper level (F18).

SERIAL

NUMBERS

*

ZERO CALIBRATION

VALUE

SPAN CALIBRATION VALUE

FI-80:

Load cell 1:

List any additional cells connected:

CAL 0:

CAL 1:

Table 6-1: Calibration Value Table

* Calibration is only valid for exact system calibrated. If any items are changed or removed, the

system must be recalibrated together or significant errors may result.


* Record the new values after each recalibration.