Welch Allyn Connex Vital Signs Monitor, Upgrade Sheet, VSM 6000, Braun Pro - Installation Guide User Manual
Read this first, Welch allyn technical support

READ THIS FIRST!
The Braun PRO 4000 Dock works only with firmware version 1.5 or higher on VSM 6000
Series monitors. Older firmware on these monitors must be upgraded before the dock can be
installed.
If your monitor part number is in the following list, you do not need to upgrade the monitor
firmware. Install the dock using the instructions on the installation sheet that accompanies
the dock or mounting kit.
If your monitor part number is not in the list, you must check the firmware version and then, if
indicated, upgrade the firmware before installing the dock. This document shows how to
check the firmware version on the monitor or through the Welch Allyn Service Tool. Choose
the option that you prefer.
Part numbers of compatible monitors:
63XXXE
64XXXE
65XXXE
63MXXE
64MXXE
65MXXE
63NXXE
64NXXE
65NXXE
63XTXE
64XTXE
65XTXE
63XTPE
64XTPE
65XTPE
63MTXE
64MTXE
65MTXE
63NTXE
64NTXE
65NTXE
63MTPE
64MTPE
65MTPE
63NTPE
64NTPE
65NTPE
63XXPE
64XXPE
65XXPE
63NXPE
64NXPE
65NXPE
63MXPE
64MXPE
65MXPE
Mat. 716726 Ver. B
8.
In the Device information tab, check the Host Controller firmware version.
•
If the version is not 1.5 or higher: You must upgrade to the latest version.
a.
Upgrade the firmware by using the Welch Allyn Service Tool. For instructions,
refer to the service-tool installation guide or the Help file in the service tool
(Help > Service Tool Help).
b.
Install the dock using the instructions on the installation sheet that accompanies
the dock or mounting kit.
•
If the version is 1.5 or higher: The monitor is compatible with the dock. Install the
dock using the instructions on the installation sheet that accompanies the dock or
mounting kit.
For more information about dock functionality, refer to the version of Welch Allyn Vital
Signs Monitor 6000 Series Directions for Use that applies to your firmware version. You
can download the directions for use from http://www.welchallyn.com.
Welch Allyn Technical Support
USA +1 800 535 6663
+1 315 685 4560
Japan
Latin America
+81 3 6383 0852
+1 305 669 9003
Australia +61 2 9638 3000
Malaysia +603 7875 3341
Canada +1 800 561 8797
Netherlands +31 202 061 360
China +86 21 6327 9631
Singapore +65 6419 8100
European Call Center +353 46 90 67790
South Africa +27 11 777 7555
France +33 155 69 58 49
Spain +34 917 499 357
Germany +49 695 098 5132
Sweden +46 85 853 6551
Italy +39 026 968 2425
United Kingdom +44 207 365 6780
4
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220 USA
www.welchallyn.com