Best practices

Welch Allyn Network installation

Critical for Connex CS and Connex RMS

Best practice

Affected types Without best practice

Ports allowed

TCP: 281, 283

UDP: 291, 7711-7719, 44435-44436 (only if Spot LXi is used
with Connex)

Wired

Connections cannot be established.

Rendezvous Perform at least one of the following:

Allow UDP broadcasts on ports 7711-7719, or 44435-44436
(only if Spot LXi is used with Connex)

DNS name resolution for Connex servers using a locally
configured name

Configure the device with a fixed IP address of the Connex
server

Wired

Connections cannot be established from the Welch
Allyn VLAN to the server.

SSID/Radio settings

a band Only

Wireless

Loss of connection and data, patient monitor will not
connect.