beautypg.com

Performing a spirometry post test – Welch Allyn CP150 Spirometry - Quick Reference Guide User Manual

Page 9

background image

13. Touch Print to print the test report, touch Test results to preview the test report on

the display, or touch ATS Interpretation to add or edit ATS interpretations. Touch
Print patient's education to print patient help sheets. (See About the patient help

sheets for further detail.) Touch Start post test to perform post medication efforts

for the current patient, or touch Return to pre test to continue with FVC pre-test

efforts.

14. Touch Done when you have completed the pre-tests.

If the Auto Save setting is turned off, touch the Yes button and touch Save to save

the test. Select one of the following locations:

Local (internal memory)

USB mass storage device (Any tests that you save to a USB mass storage

device can be retrieved only from a CardioPerfect workstation.)

Workstation

Remote file location

Performing a spirometry post test

Complete the pre-test efforts. See Performing a new Forced Vital Capacity spirometry

test.

Note

Pre- and post-efforts must happen on the same day. The next day tests
become available for review only; you can no longer add efforts to them.

1. Touch the

(Continue saved test) button.

The Spirometry saved tests screen appears.

Quick reference guide

Spirometry tests 7