beautypg.com

Recoverable spo2 alert - escalated, Recoverable nibp alert - escalated, Battery alert – Welch Allyn VSM300 Addendum To - Quick Reference Guide User Manual

Page 2: Recoverable spo, Alert — escalated, Recoverable nibp alert — escalated

background image

810-1881-00 Rev A

Addendum to the Directions for Use

2

Welch Allyn Vital Signs Monitor

Recoverable SpO

2

Alert — Escalated

An SpO

2

equipment alert is immediately escalated to the level of a patient alarm if it occurs

when

both

of the following conditions are true:

SpO

2

monitoring has begun and the monitor has recorded an SpO

2

measurement

An SpO

2

or Pulse Rate alarm limit has been set

Recoverable NIBP Alert — Escalated

An NIBP equipment alert is escalated to the level of a patient alarm whenever two
consecutive NIBP alerts occur while the monitor is taking automatic NIBP/PR
measurements and alarms are enabled.

Battery Alert

A battery alert is described in the DFU (

"Battery Low Warning"

and

"Battery Failure" on

page 87

).