Welch Allyn Spot Vital Signs Monitor - Quick Reference Guide User Manual
Error codes

Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General
internal
malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02
Autozero failure: Check for air obstruction; limit patient movement.
C03
Inflation too rapid: Check for kinked hose or air obstruction.
C04
Excessive inflation time: Check for air leaks.
C05
Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06
Measurement outside of device’s range: Check patient condition.
E10
Cuff overpressure: Check patient condition.
Temperature Error Codes
C20
Broken/missing probe: Replace probe.
P
Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6
SpO
2
heart rate out of range.
C8
Bad sensor: Replace sensor.
E7
Internal malfunction: Contact Technical Service.
C9
10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11
Internal safety violation: Check patient; contact Technical Service.
C12
Ambient temperature out of range: Adjust room temperature or device location.
C13
Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General
internal
malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Document Outline
- Error Codes
- Error Codes
- Error Codes
- Error Codes
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Error Codes
- Error Codes
- Error Codes
- Error Codes
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card
- For detailed operating instructions, refer to the Directions for Use.
- Spot Vital Signs Quick Reference Card
- On/Off Button
- Pressure Preset Button
- Mode Button
- Next Patient/Clear Button
- Blood Pressure Start/Stop Button
- Temperature Operation
- 1. Select the appropriate temperature probe. Blue = Oral/Axillary, Red = Rectal.
- 2. Remove probe and wait for OrL to appear on the display. Apply a new probe cover.
- 3. Place probe in contact with patient tissue as directed in the manual.
- 4. When done, single beep is heard and the temperature is displayed. Return probe.
- SpO2 Operation
- Spot Vital Signs Quick Reference Card