beautypg.com

Welch Allyn KleenSpec 790 Series Cordless Illumination System - User Manual User Manual

Page 20

background image

18

Directions for use

Welch Allyn

Table 2. Immunity for all EQUIPMENT and SYSTEMS

The 790 series cordless illumination system is intended for use in the electromagnetic environment
specified below. The customer or user of the 790 series cordless illumination system should assure
that it is used in such an environment.

Immunity test

IEC 60601
test level

Compliance
level

Electromagnetic environment -
guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.

Electrical fast
transient (EFT)/
burst
IEC 61000-4-4

±2 kV mains
±1 kV I/O

±2 kV mains
Not applicable

Mains power quality should be that of
a typical commercial or hospital
environment.

Surge
IEC 61000-4-5

±1 kV differential
±2 kV common

±1 kV differential
±2 kV common

Mains power quality should be that of
a typical commercial or hospital
environment.

Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines.
IEC 61000-4-11

>95% dip in
0.5 cycle

60% dip in
5 cycles

30% dip for
25 cycles

>95% dip in
5 seconds

>95% dip in
0.5 cycle

60% dip in
5 cycles

30% dip for
25 cycles

>95% dip in
5 seconds

Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the 790
series cordless illumination system
requires continued operation during
power mains interruptions, it is
recommended that the 790 series
cordless illumination system be
powered from an uninterruptible
power supply or battery.

Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.