Cleaning and, Disinfecting, Maintenance – Welch Allyn KleenSpec 790 Series Cordless Illumination System - User Manual User Manual
Page 11: Cleaning and disinfecting
Directions for use
Cordless illumination system
9
Maintenance
Cleaning and disinfecting
Use CaviWipe
®
wipes for both cleaning and disinfecting.*
*EPA-approved intermediate-level disinfectant appropriately labeled as effective against
TB mycobacterium tuberculosis (per AAMI TIR-12, 2004, section 5.2.3)
To clean and disinfect the illuminator
1.
After using the illuminator, unplug any power cords.
2.
Clean the illuminator.
a.
Following the wipe manufacturer's instructions, wipe the entire
unit to completely remove any gross debris.
b.
Gently wipe the gold contacts on the bottom with 70% alcohol.
If debris builds up on these contacts, the charging or auxiliary
power functions might be affected.
3.
Disinfect the illuminator.
a.
Following the wipe manufacturer's instructions, thoroughly wet the
surface of the entire illuminator.
b.
Using sterile gauze soaked with 70% isopropyl alcohol, thoroughly
wet the surface of the entire illuminator. The treated surface must
remain wet for five minutes. If needed, reapply the solution to keep
the surface continuously wet during this time.
c.
Allow the surface to dry.
To clean the charging base or auxiliary power cord
1.
After using the charging base or auxiliary power cord, unplug the power
cord.
2.
Clean the component. Following the wipe manufacturer's instructions,
wipe the surface to completely remove any gross debris.
Caution Use only the cleaning and disinfection methods
described in this manual. Clean the illuminator prior to disinfection.