Inflation system change and replacement, Specifications and standards – Welch Allyn DuraShock Integrated Aneroid Sphygmomanometer - User Manual User Manual
Page 4

4
Inflation System Change and Replacement
The Welch Allyn DuraShock integrated blood pressure gauge connects effortlessly to 
the inflation system via a unique port connection. Therefore, only Welch Allyn 
integrated blood pressure cuffs can be used with the Welch Allyn DuraShock 
integrated blood pressure gauge.
1.
To remove the inflation system, simply grasp the 
gauge and pull away from the cuff. Rotating the 
gauge as it is being pulled away from the cuff will 
make removal easier.
2.
To attach a new or different size blood pressure 
cuff, simply press the gauge stem into the cuff port 
until you feel it engage. Rotating the gauge as it is 
being pressed into the cuff will make insertion 
easier.
Specifications and Standards
The Welch Allyn DuraShock integrated aneroid sphygmomanometer is accurate to 
±3 mm Hg and conforms to applicable sections of the following standards for aneroid 
sphygmomanometers:
•
American National Standard ANSI/AAMI SP9-1994, Non-automated sphyg-
momanometers.
•
European Standard EN 1060-1: 1996, Non-invasive sphygmomanometers, 
Part 1: General Requirements 
•
European Standard EN 1060-2: 1996, Part 2: Supplementary requirements 
for mechanical sphygmomanometers (excluding Section 7.4.3 for pointer 
thickness).
•
INMETRO Technical Metrological Regulation Number 24 of February 26, 
1996 (excluding Section 5.4 for pointer thickness).
This product will maintain the safety and performance characteristics specified at 
temperatures ranging from 0°C (32°F) to 46°C (115°F) at a relative humidity level not 
to exceed 85%.
Calibration Check of Aneroid Sphygmomanometer
Welch Allyn recommends that the calibration of aneroid sphygmomanometers be 
checked using the following procedure on an annual basis. 
1.
Connect gauge to a high-quality, known pressure standard
1
using a Y-connector
and the DuraShock Caliberation Connector accessor (part no. 5082-255).
2.
Pressurize gauge to slightly above 300 mm Hg.
3.
Bleed pressure down no faster than 10 mm Hg per second, stopping to check the 
pressure at 300, 250, 200, 150, 60 and 0 mm Hg.
