Directions for Use

Chapter 1 Conventions

5

Cautions

A caution statement in this manual identifies a condition or practice which, if not
corrected or discontinued immediately, could lead to equipment failure, equipment
damage, or data loss.

Caution Consult users manual for important information.

Caution This system should only be used by qualified medical personnel.

Caution

To reduce the chance of direct visual exposure to the light, set the

light intensity knob to the minimum level and plug the fiber optic bundle into the
light source before turning the power ON.

Caution When the unit is on, never disconnect the fiber optic bundle from the
light source or the accessory from the end of the fiber bundle. High intensity light
from the end of the fiber optic bundle and light source can ignite flammable
materials (i.e. drapes).

Caution Grounding reliability is achieved only when connected to hospital-use
or hospital-grade receptacles. Inspect electrical plug and cord routinely. Do not
use if damage is discovered.

Caution

PROVIDE VENTILATION TO PREVENT OVERHEATING. Do not

cover or drape the light source.

Caution

Do not use the turret as a light attenuator or with the turret

misaligned with the light port, overheating will result.

Caution

The ends of the fiber bundle get hot during use. Allow to cool

before removing.

Caution Do not operate light source without lamp module in place.

Caution If this light source is used during invasive surgery, the failure of a
component within the system, causing loss of illumination, may directly cause
risk to the patient. Therefore, it is recommended to have alternative means of
illumination available.

Caution Since the light source uses a custom lamp, always have a spare lamp
available for replacement. Use only Welch Allyn replacement lamp number
04980. Use caution because lamp may be hot. Allow the lamp to cool before
replacing.

Caution

IPXØ - Equipment not protected against the ingress of water. Do not

use or store liquids above or on the light source.