beautypg.com

Maintenance, Cleaning instructions – Welch Allyn XENON 300 LIGHT SOURCE - User Manual User Manual

Page 13

background image

3

9

Maintenance

Cleaning Instructions

1.

Turn power off and disconnect the power cord from both the wall outlet and the
Welch Allyn Xenon 300 Light Source.

2.

Wipe the external surfaces clean with a cloth dampened with mild soap and water or
use one of the following cleaning/disinfecting solutions. DO NOT IMMERSE.

70% isopropyl alcohol

10% mild bleach solution

3.

Wipe the power cord clean using the same cleaning/disinfecting solutions listed
above. Wipe dry with a clean cloth. Take care not to get the plug prongs wet. DO NOT
IMMERSE.

4.

Do not plug the unit into a wall outlet until it is thoroughly dry.

Note

Do not sterilize or immerse the Welch Allyn Xenon 300 Light Source.

Caution Clean the ends of the fiber optic bundle before each use. Any residue
(including cleaning and disinfecting solutions) will absorb high intensity light on
the surface of the fiber optic. Absorbed light will overheat the residue causing
damage to the fiber optic bundle.