Troubleshooting guide – Welch Allyn KleenSpec Vaginal Specula Illumination System - User Manual User Manual
Page 14
![background image](/manuals/641586/14/background.png)
10
Troubleshooting Guide
Problem
Possible Cause
Solution
Lamp does not
light.
Power switch is in the off
position.
Verify the power switch is in the
on position. Refer to manual.
Illuminator is not plugged
into the transformer.
Verify that the illuminator is
properly plugged into transformer
cord.
Transformer is not
plugged in.
Verify the transformer is properly
plugged into the outlet.
Faulty Lamp
Replace with a new lamp.
Loose Lamp
Verify that the lamp is secured into
the illuminator.
Wrong Lamp
Verify that the lamp is a REF
08800 Welch Allyn lamp.
Damaged Lamp
Inspect the lamp and the o-ring for
damage. Replace with REF 08800
lamp if damage is found.
Illuminator does
not stay
retained into
580 series
specula or
78600
expendable
illuminator.
Possible damage to the
illuminator retention
wings.
Inspect retention wings for
damage. Replace illuminator
REF 78800.
Illuminator does
not stay plugged
into the
transformer
cord.
Illuminator/transformer
interconnect plug is
damaged.
Inspect interconnect plug for
damage. Replace either or both
REF 73405, REF 73410 or REF
78800.
Light output
seems low.
Faulty Lamp
Replace with a new lamp.
Dirty lamp lens
Inspect lamp lens for debris and
clean.
This guide is intended to provide assistance towards determining the fault
condition of this product. Please contact the authorized Welch Allyn Service
Center listed within the product owners manual for additional assistance.