6
Cleaning Disinfection, and Sterilization
Cleaning
Wipe
Transformer (REF 73405 or REF 73410), Illuminator (REF 78800), Expendable
Illuminator (REF 78600)
Refer to Cleaning Table to identify the appropriate cleaning product.
1.
Disconnect the transformer from the wall outlet and then disconnect the
illuminator from the transformer.
2.
Disconnect illuminator from expendable illuminator and remove the
expendable illuminator from the speculum (58600/58601 only).
3.
Follow the disinfecting wipe manufacturer’s instruction and wipe cord,
illuminator, or expendable illuminator completely to remove any residue.
4.
This process is a rough clean only. Refer to disinfection or sterilization for
further processing.
Soak
Refer to Cleaning Table to identify the appropriate cleaning product.
Illuminator (REF 78800) and/or Expendable Illuminator (REF 78600)
1.
Disconnect the transformer from the wall outlet and then disconnect the
illuminator from the transformer.
Illuminator
REF 78800
Transformer
REF 73405 or
REF 73410
Expendable
Illuminator
REF 78600
Cleaning
CaviWipe
(Wipe)
Yes
Yes
Yes
Enzol (Soak)
Yes
Limited*
Yes
Metrizyme
(Soak)
Yes
Limited*
Yes
Disinfection
Caviwipe
Intermediate
Level
Limited*
Intermediate
Level
Yes
Intermediate
Level
Sterilization
Autoclave
No
No
Yes**
* Do not immerse the transformer plug. Follow instructions per this manual.
** The Expendable Illuminator’s ability to transmit light will degrade after
repeated autoclave cycles (approximately 20) and will eventually need
replacement.
Lamp must remain installed during all cleaning and disinfection
processes.