Directions for use

Welch Allyn standard laryngoscope handles

Autoclave instructions

NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are
compatible with the autoclave methods identified which are provided for facilities who wish to
autoclave after cleaning and intermediate level disinfection.

Disassembly:

Remove batteries per figure 2 and set aside.

After battery removal, select ONE of the following autoclave methods below for the
main handle and bottom cap (only):

Gravity autoclave: Follow equipment manufacturer and facility procedures in the
set-up and operation of autoclave equipment. Gravity autoclave settings are as
follows:

Temperature: 132 C (270 F)

• Exposure

time:

3

minutes

(unwrapped)

• Minimum

dry

time:

1

minute

Pre-vacuum autoclave: Follow equipment manufacturer and facility procedures in the
set-up and operation of autoclave equipment. Pre-vacuum autoclave settings are as
follows:

Temperature: 132 C (270 F)

• Exposure

time:

3

minutes

(unwrapped)

• Minimum

dry

time:

1

minute

Maintenance,
Inspection and Testing

1. Inspect each component area per figure 3 for damage or

deterioration.

WARNING: Discard any component that shows

evidence of damage or deterioration.

Contact Welch Allyn for component replacement.

2. Reassemble batteries into handle per figure 4.

3. Attach handle to a clean and disinfected test blade

assembly in known working condition. Verify that:

Blade assembly engages and locks onto handle.

Blade assembly deploys into its locked position on
handle AND lamp illuminates.

• Light

output

is

satisfactory.

If the lamp fails to light or output is low, check or
replace the batteries

Storage: Store

handle

per

facility

practice to allow device to remain

clean, dry, and ready for service.