Welch Allyn Standard laryngoscope handles - User Manual User Manual
Page 4

4
Directions for use
Welch Allyn standard laryngoscope handles
Autoclave instructions
NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are
compatible with the autoclave methods identified which are provided for facilities who wish to
autoclave after cleaning and intermediate level disinfection.
Disassembly:
Remove batteries per figure 2 and set aside.
After battery removal, select ONE of the following autoclave methods below for the
main handle and bottom cap (only):
Gravity autoclave: Follow equipment manufacturer and facility procedures in the
set-up and operation of autoclave equipment. Gravity autoclave settings are as
follows:
•
Temperature: 132 C (270 F)
• Exposure
time:
3
minutes
(unwrapped)
• Minimum
dry
time:
1
minute
Pre-vacuum autoclave: Follow equipment manufacturer and facility procedures in the
set-up and operation of autoclave equipment. Pre-vacuum autoclave settings are as
follows:
•
Temperature: 132 C (270 F)
• Exposure
time:
3
minutes
(unwrapped)
• Minimum
dry
time:
1
minute
Maintenance,
Inspection and Testing
1. Inspect each component area per figure 3 for damage or
deterioration.
WARNING: Discard any component that shows
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble batteries into handle per figure 4.
3. Attach handle to a clean and disinfected test blade
assembly in known working condition. Verify that:
•
Blade assembly engages and locks onto handle.
•
Blade assembly deploys into its locked position on
handle AND lamp illuminates.
• Light
output
is
satisfactory.
If the lamp fails to light or output is low, check or
replace the batteries
Storage: Store
handle
per
facility
practice to allow device to remain
clean, dry, and ready for service.