Reprocessing instructions – Welch Allyn Standard laryngoscope handles - User Manual User Manual
Page 2

2
Directions for use
Welch Allyn standard laryngoscope handles
Reprocessing instructions
These reprocessing instructions refer to procedures for cleaning and intermediate level
disinfection. Standard laryngoscope handles must be reprocessed prior to first use and
between each use using the following method as outlined in this document:
•
Cleaning and intermediate level disinfection
Welch Allyn has validated the above instruction as being capable of preparing these
laryngoscope handles for re-use. The user must ensure that the reprocessing as actually
performed by the user's personnel, with the user's equipment and materials, achieves the
desired result. This may require validation and routine monitoring of the user's actual process
NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are
compatible with the autoclave methods identified which are provided for facilities who wish to
autoclave after cleaning and intermediate level disinfection.
Cleaning and intermediate level disinfection instructions
Point of use: 1. Separate blade assembly from handle and place handle
into suitable containment for subsequent reprocessing
per figure 1. Do not place handle with sharp devices.
2. Prevent the handle from drying (i.e. wrap/cover in moist
germicidal wipe).
Preparation for
decontamination:
1. Select an appropriate quaternary ammonium
isopropanol based germicidal cleaner labeled suitable
for use on healthcare equipment and capable
intermediate level disinfection. Reference EPA-
registered disinfectants:
http://www.epa.gov/oppad001/chemregindex.htm
Outside of the U.S., please consult applicable
regulatory body for equivalent quaternary ammonium
isopropanol germicidal cleaner.
2. R emove batteries per figure 3.