0 rs-232 communication, 1 pushbutton keypad print, 2 usb connection – Rice Lake Waist Level Digital Physician Scales (150-10-6) - Operation and Setup Manual User Manual
Page 14: Rs-232 communication, 1 pushbutton keypad print 5.2 usb connection

10
Rice Lake Digital Physician Scale Operation Manual
5.0
RS-232 Communication
The scale comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a
computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing
Systems. Figure 3-8 on page 6 shows where the RS-232 connection is.
The RS-232 parameters are 9600 baud (selectable in the programming mode), 8 data bits, 1 stop bit, no parity
and no handshaking.
There are three methods of communication:
•
Pushbutton keypad print
•
Standard remote protocol
•
Escape protocol
5.1
Pushbutton Keypad Print
With a stable, in-range weight, press and hold the
LB/KG/Print
key for at least three seconds, or until the scale
displays
. Note that if the scale does not beep after five seconds, release the button as the weight was either
in motion or out of range.
•
If displaying weight and not BMI, the scale will send out the following 21 character string:
xxxxxxxxx
Where:
xxxxxxxxx is the weight with decimal point and " - " sign, if negative uu is the unit (lb or kg).
mmmmm is the mode (gross or net)
Examples:
-10 Lb net =
10 Lb gross =
•
In BMI mode (displaying the BMI value), the scale will send out the following data:
GROSS WEIGHT 215.0 LB
TARE WEIGHT 0.0 LB
NET WEIGHT 215.0 LB
PATIENT HEIGHT 6-01.0 FT
PATIENT BMI 28.4
5.2
USB Connection
The Rice Lake Digital Physician Scale has the capability of connecting to a PC using a USB connection, and a
USB cable (not included), as shown in Figure 5-1.
USB Connection
Figure 5-1. USB Connection Port