beautypg.com

Rice Lake Scout Pro Series Portable, Ohaus - Calibration Guide User Manual

Calibration guide scout

background image

Calibration Guide

Scout

Pro

Ohaus Corporation www.ohaus.com

i

Scout

Pro

Calibration Guide

PREPARATION

The Scout Pro has two calibration processes: Span and Linearity.
Follow the prompts for calibration mass values.

Ensure that the balance is level, stable, and undisturbed during calibration.

(Otherwise the process could fail.)

Allow balance to warm up for five minutes after stabilizing to room temperature.

Use suitable calibration masses (available from Ohaus).

If the Calibration Menu is locked, the scale may be locked with a software “lock switch.”

Turn the lock switch off.

(See Instruction Manual.)

Start with the balance OFF. After calibration, the balance returns to weigh mode.

Note: To cancel calibration, long-press:

SPAN CALIBRATION

5

(Release button)

1

(Acquiring zero value)

(Place mass on platform)

1

(example)

(Remove mass)