RL2200 Load Cell Mounting Kit Installation
3.0 Load Cell Wiring
1. Route the load cell cables so they will not be damaged or cut. Cable should not be
routed near heat sources greater than 150 °F. Do not shorten any load cell cable.
The load cell is temperature compensated with the supplied length of cable. Cut-
ting the cable will affect temperature compensation. Coil and protect excess
cable so it will not be mechanically damaged or be sitting in water.
2. Provide a drip loop in all cables so that water or other liquids will not run directly
down the cables onto either the load cells or the junction box. Attach load cell
cable to the dead structure, not the vessel.
3. If conduit protection is necessary against mechanical or rodent damage to the
load cell cables, use flexible conduit and conduit adapters at the load cells.
4. Connect cables for load cells to the summing board in the junction box according
to the guide shown below and the labels on the terminal strips of the junction
box. To verify the wiring scheme, see the certification shipped with each load
cell.
5. For better performance, use positive and negative remote sense lines if the wiring
running from the junction box to the indicator is longer than 25 feet.
Drip Loop
Load Cell Wire Colors
Wire Color
Function
Negative Reading
Positive Reading
Red
+EXC
+EXC
Black
-EXC
-EXC
Green
+SIG
-SIG
White
-SIG
+SIG
Gray or Bare
Shield
Shield