2 checks with multimeter, 1 check zero output signal of load cell, 2 check the strain gauge bridge – Rice Lake PR6201 Zinc-Plated Steel Star Mount Kits User Manual
Page 24: 3 check insulation impedance of load cell, 4 check insulation impedance of connecting cable, Compression load cell pr 6201 installation manual

Compression Load Cell PR 6201
Installation Manual
EN-22
Sartorius
7.2
Checks with Multimeter
7.2.1
Check Zero Output Signal of Load Cell
• Unload load cell.
• Disconnect the load cell outputs.
• Check whether the output voltage without load is within the limits.
Type
Zero output signal
L
0 mV ±0,02 mV/V
D1/N/C3
0 mV ±0,01 mV/V
(at PR 6201/54..
0 mV ±0,02 mV/V)
LA
3,2…4 mA
GAIN connected to +Supply Voltage 24 ±4 V, see chapter 5.2.2.1
7.2.2
Check the Strain Gauge Bridge
Note:
Not for load cells type LA.
• Do not exceed the test voltage.
• Check whether the values of the resistors are within the permissible limits.
Maximum test voltage
Standard industrial
32 V DC
Intrinsically safe version (PR …/..E)
25 V DC
Type
Input impedance
(red core, blue core)
Output impedance
(green core, grey core)
L
650
Ω +50 Ω
610
Ω ±3 Ω
D1/N
650
Ω ±6 Ω
610
Ω ±1 Ω
C3
650
Ω ±6 Ω
610
Ω ±0,5 Ω
7.2.3
Check Insulation Impedance of Load Cell
Note:
Not for load cells type LA.
• Never apply the test voltage between the cores of the load cell cable (danger of destroying the load
cell).
• Insulate the load cell cores.
maximum test voltage
standard industrial
100 V DC
Intrinsically safe version (PR …/..E)
500 V AC
insulation impedance
core - housing
>5000 M
Ω
core - screen
>5000 M
Ω
screen - housing
>5000 M
Ω
7.2.4
Check Insulation Impedance of Connecting Cable
• Disconnect connecting cable and load cells.
• Insulate the cores of the connecting cable.
insulation impedance
core - core
≥600 MΩ × km
core - screen
≥600 MΩ × km