0 rs-232 communication, 1 pushbutton keypad print, 2 usb connection – Rice Lake Portable Bariatric Wheelchair Scale Single Ramp (350-5) - Rice Lake Wheelchair Scale Operation Manual User Manual
Page 16
12
Rice Lake Wheelchair Scale Operation Manual
6.0
RS-232 Communication
The scale comes with an RS-232 port which enables weight data to be transmitted to other equipment, such as a
computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing
Systems. Figure 6-1 shows the RS-232 connection location.
The RS-232 parameters are 9600 baud (selectable in the programming mode), 8 data bits, 1 stop bit, no parity
and no handshaking.
There are three methods of communication:
•
Pushbutton keypad print
•
Standard remote protocol
•
Escape protocol
6.1
Pushbutton Keypad Print
With a stable, in-range weight, press and hold the
Kg-Lb/Print
key for at least three seconds, or until the scale
emits two quick beeps. Note that if the scale does not beep after five seconds, then release the button as the
weight was either in motion, or out of range.
•
If displaying weight and not BMI, the scale will send out the following 21 character string:
xxxxxxxxx
Where:
xxxxxxxxx is the weight with decimal point and " - " sign, if negative, uu is the unit (lb or kg).
mmmmm is the mode (gross or net)
Examples:
-10 Lb net =
10 Lb gross =
•
In BMI mode (displaying the BMI value), the scale will send out the following data:
GROSS WEIGHT 215.0 LB
TARE WEIGHT 0.0 LB
NET WEIGHT 215.0 LB
PATIENT HEIGHT 6-01.0 FT
PATIENT BMI 28.4
6.2
USB Connection
The Rice Lake Wheelchair Scale has the capability of connecting to a PC using a USB connection and a USB
cable (not included). That connection location is shown in Figure 6-1.
USB Connection
RS-232 Connection
Figure 6-1. USB Connection Port and RS-232 Connection Port