OS-10/25/45 Hanging Scale
Operator’s Manual
Hold Function
1. Press
HOLD
, when weight stabilizes display will lock and arrow will be steady.
2. Press
HOLD
to release locked weight.
Adjusting Auto-Off
1. Press
o
POWER ZERO
to power on.
2. Press and hold
UNITS
Units
. When AOFF displays, release.
3. Press
HOLD
to select auto-off time (OFF, 120, 180, 240, 300 seconds).
4. Press
UNITS
Units
to return normal weighing mode.
Calibration
Calibrate using the following process:
1. Press and hold
o
POWER ZERO
.
2. Press
UNITS
Units
three times, ZERO COUNT displays.
3. Press
o
POWER ZERO
, zero displays.
4. Press
UNITS
Units
to select calibration unit.
5. Load appropriate amount of weight on scale (See Table 3). Display shows span count.
6. Press
o
POWER ZERO
, CAL displays, then the weight.
7. Remove the weight, re-power the scale and calibration is complete.
Model
kg
lb
Table 3. Calibration Weights
OS10
10kg
20lb
OS25
20kg
40lb
OS45
20kg
40lb