0 scale configuration – Rice Lake Bariatric Handrail (250-10-2) - Rice Lake Digital Handrail Scale Technical Instructions User Manual
Page 12
8
Digital Handrail Scale Technical Manual
5.0
Scale Configuration
Options and parameter setup are done through the scale configuration section and is used for setting values,
parameters, and options that are essential for the functioning of the scale. Access to the setup switch is located
under the tilt stand cover. Use a Phillips head screwdriver to remove the four screws holding the cover in place.
(shown in Figure 5-1— left photo).
Remove four screws to access setup switch
Setup switch location
B a t t e r y
compartment
Figure 5-1. Setup Switch Location
Use the following steps to enter into configuration mode.
1. Make sure the scale is turned off.
2. Turn the scale on by simultaneously pressing the
On/Off
key and the
ENTER
key.
3. Continue to hold both keys until
Id
appears. The unit cycles through its startup function and continues to
display the software version.
4. Access the recessed setup switch located in the back of the indicator. You can use a small paper clip,
small screwdriver or other similar object to press the setup switch.
5. Once the setup switch is pressed,
PROG
appears on the display.
6. The scale can be configured using a series of menus accessed through the front panel when the scale is in
setup mode.
DEF
CALIB
PROG
DONE
Figure 5-2. Top Level Menu
7. Press the
BMI
key to advance to the desired menu.
8. Press the
ENTER
key and advance in the manual to the related menu selection for further instructions.