ETS-Lindgren FP5080 Isotropic Electric Field Probe (Archived) User Manual

Page 34

background image

NiCd Battery Fast Charger

© ETS-LINDGREN L.P. – December 2004

REV B - PN H-600086

PROBE OUTPUT


Command HI-4450/FP2080/FP4080/FP5080

Response


B

Bxx.xx, where xx.xx is the battery voltage.


D1

Dxx.xxuuu, the short form output.

xx.xx is the reading. The position of the

decimal point depends upon the range setting of
the probe.

uuu

=

units

_V_ = V/m, mW2 = mW/cm², _V2 = [V/m]²

(underscore

indicates

a

space

character)


D2

Dxx.xxuuurrrobaaat, the long form output.

xx.xx = the reading, as described for D1.

uuu = units, as describe for D1.
rrr = recorder out value (A 3-digit ASCII number
from 0 to 255).
o = over range indicator ("N" = ok, "O" = over
range).

b = battery status ("N" = safe operating level,
"W" = warning level, "F" = fail level).

aaa = axis enable ("E" = enabled).

t = terminating carriage return.


Rx

Rx, where x is the range.

x = "" returns the range currently in use

x = 1, 2, 3, 4 enables the selected range

x = N sets the probe to the next (higher) range.


TF

Txxx, where xxx is temperature in ° Fahrenheit.


TC

Txxx, where xxx is temperature in ° Centigrade.


Vtx

Vxx[xx...], where xx is table data in ASCII hex.