ETS-Lindgren FP5080 Isotropic Electric Field Probe (Archived) User Manual
Page 34

NiCd Battery Fast Charger
© ETS-LINDGREN L.P. – December 2004
REV B - PN H-600086
PROBE OUTPUT
Command HI-4450/FP2080/FP4080/FP5080
Response
B
Bxx.xx, where xx.xx is the battery voltage.
D1
Dxx.xxuuu, the short form output.
xx.xx is the reading. The position of the
decimal point depends upon the range setting of
the probe.
uuu
=
units
_V_ = V/m, mW2 = mW/cm², _V2 = [V/m]²
(underscore
indicates
a
space
character)
D2
Dxx.xxuuurrrobaaat, the long form output.
xx.xx = the reading, as described for D1.
uuu = units, as describe for D1.
rrr = recorder out value (A 3-digit ASCII number
from 0 to 255).
o = over range indicator ("N" = ok, "O" = over
range).
b = battery status ("N" = safe operating level,
"W" = warning level, "F" = fail level).
aaa = axis enable ("E" = enabled).
t = terminating carriage return.
Rx
Rx, where x is the range.
x = "" returns the range currently in use
x = 1, 2, 3, 4 enables the selected range
x = N sets the probe to the next (higher) range.
TF
Txxx, where xxx is temperature in ° Fahrenheit.
TC
Txxx, where xxx is temperature in ° Centigrade.
Vtx
Vxx[xx...], where xx is table data in ASCII hex.