When it completes the command, the probe responds with a
string consisting of

1) S start character (":")
2) The command letter, followed by
3) Data, if required, and terminated with
4) A carriage return

If the command does not require the probe to return any data,
the probe simply responds with the command letter and a
carriage return. If an error occurs, the probe responds with an
error code, as detailed in Appendix A.

COMMANDS

Command Description
B

Read battery voltage.


Cx

Set baud rate.

x = 1 sets rate to 2400 baud

x = 2 sets rate to 9600 baud
New baud rate does not take effect until the next
power-up.


Dx

Read probe data.

x = 1 enables short form output

x = 2 enables long form output


Lx

Load table data. x = ASCII hex data


Rx

Set range. x = 1, 2, 3, 4 or N (next range)


Sx

Sleep timer. x = number of seconds to wait for a
command before putting the probe into the sleep
mode.


Tx

Read Temperature. x = C or F

Archived 4/02/10