HI-3520 Manual

Page 19

the alarm just starts to indicate. Repeat this

adjustment as required.

10.0

This completes the RF calibration. Replace the

battery compartment cover. Attach a new

calibration sticker with the current calibration

date.

NOTE:

Do not place any stickers or labels that contain
metal on the bottom surface of the unit or the
lower portions (i.e., approximately 1 ½ inches
up from the bottom surface) of the front and
both sides of the unit. This could seriously
affect the accuracy of the unit. Paper or plastic
stickers are acceptable.